Pular para o conteúdo principal
GE Healthcare · MRExcite2252310

Intermittent DD fault due to defective cable.

Symptom

Intermittent Driver Module Errors DD Direct Drive Open

Error message

2252310 DD Driver Fault Detected - Open Circuit. Source of Fault is DD (Direct Drive inside the Body Hybrid Assembly). See Long Description Extended Error: This usually indicates that a cable somewhere in the circuit path is not connected or one of the load diodes has failed open. Inspect the failing path. Use TR DD MC System Path diagnostic to troubleshoot. In HD/HDx Systems, BD1 is the Dynamic Disable Board at Front of Body Coil BD2 is the Dynamic Disable Board at Front of Body Coil BD3 is the Dynamic Disable Board at Back of Body Coil BD4 is the Dynamic Disable Board at Back of Body Coil In 20.0 and later systems, try running the TR DD System Path diagnostic to troubleshoot.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required