GE Healthcare · MRExcite2253519369842228010
AP ERROR and reconstruction failure due to slave AP board.
Symptom
No Images Reconstruct even though the Acquisition Completes See Long Description TPS resets and the system scans all the way through but no images reconstruct. There are AP errors in the log. This is not intermittent. Happens on every scan.
Error message
2253519 The system found bad BAM memory in the AP board(s). The bad spot starts at xxx Meg from the beginning of AP BAM memory.-> see long description When executing ap diagnostics the following key error occurs. Error: 2253519 Host: LAAD-GEEXMR Process: diagCoord File: apBld.c Line: 645 Diagnostic: AP 2 Memory Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d258 Time: 36984 Object: APS Test Task The system found bad BAM memory in the AP board(s). The bad spot starts at 125 Meg from the beginning of AP BAM memory.-> Extended Error : If the bad spot is below 852 Meg then try replacing AP board 1 If the bad spot is above 852 Meg try replacing AP board 2. 2228010 AP Error
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2325947-7 EXCITE REFLEX 200 Catalog M300RE For Excite 1 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 1 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.