GE Healthcare · MRExcite2224854

Intermittent artifacts - grainy images with 2224854 SRI communication errors caused by old style Multicoil asm.

Symptom

MRIQ Images are Grainy, Noisy Low SNR See Long Description See example images here.

Error message

2224854 SRI 8 Channel Switch Comm Error. SRI Received Packet from the 8 Channel Switch which did not Include all Parameters. Wed Apr 14 08:32:03 2004 Error: 2224854 Host: exmr Process: NSP File: SCP:SRI Line: 350 SRI-8 Channel Switch Comm Error. SRI received packet from the 8 Channel Switch which did not include all parameters. If "Actual Cnt" is 0, fiberoptic cables between SRI and 8 Channel Switch may be disconnected. Cmd: 4 Expect Cnt: 5 Actual Cnt: 0

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
5115145 Multi coil enclosure assembly_Excite3_1.5T Empty multicoil assembly - no preamps. Not for 11.x systems. Multi coil enclosure assembly_Excite3_1.5T Part is applicable to 1.5T HD and HDxt LPCA (upgrades).
Area
MR

Full cause & step-by-step resolution