GE Healthcare · X-ray VascularInnova50111
174289 Detector lift won't move up or down.
Symptom
The Detector or II Lift does not drive up or down. When attempting to drive the II/Detector lift from the TSUI (smart handle or smart box) It does not drive up or down.n.
Error message
error Code = 50111, hw-status=0003 FRT-Lift error Text = POS SMA: Encoder, potentiometer failure, debug Text = pos-error=4ABC, hw-status=0003, usr-status=2000 , recovery Class = DO, major Function /= LC Gantry System impact: X-ray aborted & Gantry lock-up. Displayed message : None. 0x4ABC 0x0000 HW_FRT_L 0x0001 HW_FRT_PVT 0x0002 HW_FRT_ARC 0x0003 HW_FRT_II-Lift 0x0000 HW_FRT_L 0x0001 HW_FRT_PVT 0x0002 HW_FRT_ARC 0x0003 HW_FRT_Iift System impact: X-ray aborted & Gantry lock-up. Displayed message : None.
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Component
- Lift Encoder Cable - 2308670 Lift Encoder Cable - 2308670 Lift Encoder Cable - 2308670. This cable must be replaced every 5 years per the PM specification. Part #: 2308670
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.