GE Healthcare · MRSignaExcite222573922255142225738
Power monitor fault due to SSM
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225739 RF Interface (RFI) High Voltage Relay Fault. Thu Mar 7 16:22:42 2002 2225739 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3066 RF Interface (RFI) High Voltage Relay Fault. Possible causes: RF Monitor Cable to System Support Module (SSM) is disconnected or defective. Note: This fault may occur in conjunction with a Power Monitor Fault or when power cycling the SSM.> 2225514 SSM fault. +15 volt switching power supply low 2225514 Host: lxmr2 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6020 SSM fault.^A +15 volt switching power supply low. 2225738 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Thu Mar 7 16:22:41 2002 2225738 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 40 pulse width: 33 duty cycle: 03 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when power cycling SSM.>
Applies to
- Models
- Signa, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.