GE Healthcare · MRSignaExcite222573922255142225738

Power monitor fault due to SSM

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225739 RF Interface (RFI) High Voltage Relay Fault. Thu Mar 7 16:22:42 2002 2225739 2 lx-mr NSP HPC_PORT_ID:spi_rf_mon 3066 RF Interface (RFI) High Voltage Relay Fault. Possible causes: RF Monitor Cable to System Support Module (SSM) is disconnected or defective. Note: This fault may occur in conjunction with a Power Monitor Fault or when power cycling the SSM.>

2225514 SSM fault. +15 volt switching power supply low 2225514 Host: lxmr2 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6020 SSM fault.^A +15 volt switching power supply low.

2225738 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Thu Mar 7 16:22:41 2002 2225738 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 40 pulse width: 33 duty cycle: 03 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when power cycling SSM.>

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.