GE Healthcare · MRExcite1100911046
Unable to transfer images to the destination due to disk at the destination was too full.
Symptom
Customer is Unable to Transfer Images to the AWW Customer is Unable to Transfer Images to the AWW.
Error message
11009 Association was aborted with IMG1 Fri Jul 14 14:16:08 2000 Error: 11009 Host: lx-mr Process: dcserver File: dcspush.c Line: 1437 Association was aborted with IMG1 11046 STORE response received with refused status (0xA700) - Aborting! Fri Jul 14 14:16:08 2000 Error: 11046 Host: lx-mr Process: dcserver File: dcspush.c Line: 1302 STORE response received with refused status (0xA700) - Aborting!
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution