Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedVCT26013279650089

'DAS gain calibration has not been run since converter boards were changed' while running fast cals.

Symptom

Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.

Error message

Error: 260132796 The Following Digital Interface Boards Are Missing/Faulty. See Long Description for complete error Mon Feb 6 09:42:29 2006 Error: 260132796 Host: orp Process: orp_subsys File: IfbConfig.cxx Line: 96 Mon Feb 6 09:42:28 2006 Host : Das Ermes # : 260132796 Exception Class : Abort Severity : Secondary File : IfbConfig.cxx Line# : 96 Function : No System Function Reported Scan Type : Static Scan: 50089/1/1 The Following Digital Interface Boards Are Missing/Faulty. Digital Interface Boards: 19

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.