GE Healthcare · MRExcite2248131221852722185372248102101512248121
Unable to scan
CWA Unable to scan.
Symptom
Unable to scan
Error message
2248131 APS DC Send DRF info error -1 Tue May 4 20:27:03 2004 Error: 2248131 Host: lx-mr Process: NSP File: APS:DacqControl::processPkt Line: 536 APS DC send DRF info error -1 Software error, please perform a TPS Reset. 2218527 Scanning has Stopped. NO MrMail Available Mon May 5 21:32:39 2008 Error: 2218527 Host: CMUNMR1 Process: scn File: MsaScanOpsScrn.m Line: 916 Scanning has stopped. NO MrMail available. 2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248 2248102 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset. See Long Description 2248102 qstmr1 NSP APS:processPkt 896 Received invalid cm packet 10151, scan state = 6 Software error, please perform a TPS Reset. 2248121 APS Send DRF X cmd Message Error (no response) Tue May 4 20:27:03 2004 Error: 2248121 Host: lx-mr Process: NSP File: APS:DrfDriver::sendCmdMsg Line: 647 APS Send DRF 1 cmd message error (no response) DRF is not running, please perform a TPS Reset.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.