GE Healthcare · X-ray VascularInnova1070210

No table longitudinal motion:Innova 4100:IQ+ or PRO Table

Symptom

IQ+ or PRO Table : No table longitudinal motion IQ+ or PRO Table : No table longitudinal motion. See long description for further details. Table cannot move on longitudinal axis. Other axis are good : Lateral, Lift and Tilt. Table recalibration does not help. The only way to unlock longitudinal motion is to press the 2 push buttons on DC-DC board and keep them pressed. In this case, longitudinal brakes are released thus allowing to manually move the table top. The motorized motion still does not work.

Error message

IQ+ or PRO Table errorcode 1070210 : Motor power wire not detected for longitudinal axis 1070210 : Motor power wire not detected for longitudinal axis Checked longitudinal Servo Node leds status and compare to other SN to find differences : DS14 is on only for Long. axis but it's normal. This limit swith is only for Long and is pressed for Tilt angle = 0 DS14 Home Limit Switch Green (Led ON)Home Limit Switch not pressed (Led OFF)Home Limit Switch is pressed N/A DS19 & DS20 stays off while it should be on in standby : LED ON LED OFF DS19 Velocity Encoder A – differential driver Green Vel. Encoder A= HIGH Vel. Encoder A/ = LOW Motor moving DS20 Velocity Encoder B – differential drive Green Vel. Encoder B/ = HIGH Vel. Encoder B = LOW Vel. Encoder B = HIGH Vel. Encoder B/ = LOW Motor moving

Applies to

Models
Innova
Product
X Innova 4100 X Do not use---------4100 systems 4100 systems
Software
$ IQ+ or PRO table ALLIANCE M3 ALL_2.2.16 $ IQ+ or PRO table ALLIANCE M3 ALL_2.2.16 $ IQ+ or PRO table ALLIANCE M3 ALL_2.2.16 M3 release for Alliance program (IQ table (tilting table) introduction on Harmony configuration).
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.