RF Amplifier Peak RF Device Current Overload Fault Intermittently Aborts Scans Only on Medium to Large Lumbar Patients U…
RF Amplifier Peak RF Device Current Overload Fault Intermittently Aborts Scans Only on Medium to Large Lumbar Patients Using High TG due to Body Hybrid Splitter.
Symptom
RF Amplifier Peak RF Device Current Overload Fault Intermittently Aborts Scans Only on Medium to Large Lumbar Patients Using High TG. See Long Description Scans intermittently abort only on medium to large patients when scanning lumbar with high TGs. Failure occurs 2-3 times per week. Errors are logged on failure as shown below. 2254367 "RF Amplifier Peak RF Device Current Overload Fault. The RF Amplifier detected a peak current overload condition in one or more RF devices. Upper spec: %d A. ->" 2254055 "Recovery in progress, see log."
Error message
2254055 Recovery in progress, see log 2254367 RF Amplifier Peak RF Device Current Overload Fault. The RF Amplifier detected a peak current overload condition in one or more RF devices. See Long Description Upper spec: %d A. -> Extended Error: Verify proper PSD is used. Check center frequency setting in Manual Prescan; this fault could result if the frequency is off by a large amount. RF Amplifier may be operating out of specification. If fault persists, replace the RF Amplifier
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 14.0 For Hino Excite HDe 14.0 systems only.
- Component
- 5219884-2 NMR Teledyne Hybrid Splitter Kit Upgraded from 46-321312G3 Upgraded from 46-321312G3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.