"This sequence cannot be reviewed" displayed on live In Room monitor.

Symptom

Review of this sequence unavailable; try other sequence or call service" When attempting to review a sequence on DL console, following message "This sequence cannot be reviewed. Try another sequence or Call Service" is displayed on the live In Room monitor. Review of this sequence unavailable; try other sequence or call service" When attempting to review a sequence on DL console, following message "This sequence cannot be reviewed. Try another sequence or Call Service" is displayed on the live In Room monitor. error 2800007

Error message

2800007 Review of a sequence is unavailable, due to data inconsistency 2800007 "Review of a sequence is unavailable, due to data inconsistency" WR XRSubSysMgrDL review File_Corruption_Data_base 1 0 calling theViewCineAuto.pipePause ViewMainAuto.cxx 630

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
DL IMAGE DISKS DL IMAGE DISKS see long DL 36GB IMAGE DISKS 5144906 SCSI drive DL1 Image Disks Kit includes (2) 36GB image drives (ST336705LW), rack, fans, and cage. If no stock check part 2386702, also check SN4568B for alternate parts. P.Liebau 05/2009 DL 72GB IMAGE DISKS 5129897 *** OLC 2nd Opinion Part *** Kit contains (2) 72GB drives along with the mounting bracket, SCSI cable and terminator, for DL5+ and DL6 IP image disks. 1 FRU replaces both image disks. Contact your local OLE for approval. Replaced by new part 5248652 but that does not include the SCSI cable. P.Liebau 02/2009 5199031 See SN4844 for kit contents. Replaces 2332254 and 2386700 for DL2 to DL6. P.Liebau 11/2010 5248652 Kit contains (2) 72GB image disks and SCSI terminator, does NOT include SCSI cable which is part of DL internal cable set 5129896. P.Liebau 02/2009
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.