GE Healthcare · X-ray VascularInnovaAdvantx1050168

Gantry moving slow due to DETECTOR collision sensor errors.

Symptom

Gantry moving slow due to DETECTOR collision sensor errors. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. Gantry moving slow due to detector collision errors. This issue is caused by a collision detected on the Detector side of the gantry. Typically indicated by HW codes 0004 for the detector faceplate or 0008 for the detector bladder. System impact: Message displayed on monitors and BIP on TSSC every 30 seconds. Full gantry motion. Displayed message : "Collision detection failure, continue with care; Otherwise call Service" The Hardware code tells whether the problem is located on the Detector or on the Tube side: HW=4 for the Detector Faceplate HW=8 is the Detector Bumper and HW=10 for the Tube . Check the anti collision sensors on the Detector. Touch the collision sensor margin on one side to re-balance the collision sensors. Touch the collision sensor margin on both sides to clear the collision error. If the problem persists (i.e., the error is not cleared), replace the corresponding ANTI-COLLISION COVER

Error message

error Code = 1050168, either be HW 0004 or H:0x8 Text = FRT II BUMP, debug Text = - LIFT COLLISION (T:0x4A12 H:0x8 U:0x0000)., recovery Class = WR, major Function /= GantryAxis System impact: Half speed until system level1 collision is cleared. Displayed message : "Detector collision occurred. Please reverse motion to continue.". Will stop Bolus chase run too System impact: Half speed until system level1 collision is cleared. Displayed message : "Detector collision occurred. Please reverse motion to continue.". feedback, Says HW 004 but has a mismatch as the actual HW is H:0x8. this should either be HW 0004 or H:0x8 Will stop Bolus chase run too

Applies to

Models
Innova, Advantx
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.