GE Healthcare · X-ray VascularInnovaAdvantx1050613
Gantry moving at half speed with no errors.
Symptom
Gantry moving at half speed with no errors. Gantry moves at half speed, but detector lift speed is correct and caqpactor sense is working correctly.
Error message
1050613 PIB XXXV Monitoring/Overtemp 1050613: Root Cause: Component(s) - Martek module problem in MPDU This error is due to an error found by one of the PIB on the Positioner Modular power units. Recommended Action: If the error name is PIB 180V MONITORING, check the 180V Module, the backplane and the PIB input. If the error name is PIB 48V MONITORING, check the 48V Module, the backplane and the PIB input. If the error name is PIB 180V OVERTEMP, check the 180V Module for overtemperature conditions. If the error name is MARTEK 48V OVERTEMP, check the 48V Module for overtemperature conditions. Replace parts found faulty.
Applies to
- Models
- Innova, Advantx
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Omega IV Cardiac table Omega IV Cardiac table Omega IV cardiac table Omega V Table Omega V Table Omega V Table Innova 2100–IQ Case Innova 2100–IQ System can be supplied with an OMEGA IV Compact patient table, an OMEGA V long patient table, which includes a table side control with Contour Filter, or a InnovaIQ patient Table. Innova 3100/3100-IQ and 4100/4100-IQ Case Innova 3100/4100 Systems can be supplied with an OMEGA V Long patient table or InnovaIQ patient Table, which includes a table side control with Contour Filter. Illustration 2-1: Omega patient table (with TSSC)
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.