Xray unavailable message on live display with fluoro and record modes with footswitch depressed.
Symptom
Xray unavailable message on live display with fluoro and record modes with footswitch depressed. System must be shutdown and rebooted to recover as a reset will not recover. May be able to produce radiation Fluoro and Record, X-ray Tone and indicator are on but "X-ray Unavailable" or "X-ray Abort" Message on Screen and no Image on Monitor. This could also cause a Database Inconsistency error because the RTAC is sending the acquired image data to the DL but the PPC bd in the DL is not receivin Xray unavailable message on live display with fluoro and record modes with footswitch depressed. System must be shutdown and rebooted to recover as a reset will not recover. May be able to produce radiation Fluoro and Record, X-ray Tone and indicator are on but "X-ray Unavailable" or "X-ray Abort" Message on Screen and no Image on Monitor. This could also cause a Database Inconsistency error because the RTAC is sending the acquired image data to the DL but the PPC bd in the DL is not receiving because the FCIB in DL has locked up or failed. Hence no image on Monitor and Data was not written to disk which could generate a Database Inconsistency.
Error message
3602032, heater 1 board defect, DIAG0037 - Jedi #20005 - Heater Diagnostics The purpose of this test is to drive the heater inverter(s) on both filaments and all the tubes connected to the Generator in order to identify a faulty heater FRU or a wrong connection between the heater board, the HV Tank and the tube. No manual interaction with the use of the Diagnostic screen on system console. However to improve the diagnostic in case of failure, the operator should make some manipulation (like remove HV connector and replace it by HV tools or short circuits some connection). jedi errorcode:1215h heater1:filament selection error dtarelated to error
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.