GE Healthcare · MRExcite2000023522000023722249372170626922494241094564028

Unable to ping the Term Server

Symptom

Cannot Boot Applications with Host Monitor Timeout Error

Error message

200002352 A HOST_SHUTDOWN was initiated by the death of NSP.

200002372 The process NSP took too long to start up. Thu Jan 27 08:56:01 2005 Error: 200002372 Host: lx-mr Process: hostMonitor File: ssProcStart.c Line: 933 The process NSP took too long to start up.

2249372 Communication problem with the host. Remote command execution (rcmd) failed. See long description for complete error. Fri Apr 1 16:52:38 2005 Error: 2249372 Host: lx-mr Process: diagCoord File: mgdPosixSync.c Line: 1215 Task ID: 0x3ff00c0 Time: 1706269 Object: APS Communication problem with the host. Remote command execution (rcmd) failed. Host: host; RSH Port: 514; Local User Name: genesis Remote User Name: genesis; Remote Command: getHostTime; fd2p pointer: 0->

2249424 Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. see long description 1094564028 0 1 Tue Sep 7 13:33:48 2004 2249424 2 CHMMR diagCoord mgdPosixSync.c 1081 Task ID: 0x3ff00c0 Time: 32822 Object: APS Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. Host: host; SNTP Timeout: 100; Errno: 0x710005-> EN 298 SR 298

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.