The Home Flag on the Longitudinal Drive Belt was not passing through the Home Flag Detect Optical Coupler.

Symptom

The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered See Long Description The carriage fails to advance to meet the table when it is raised, and fails to retract when it is undocked or lowered. But the customer can manually attach the cradle to the carriage and then table operation is normal driving in and out and they can scan okay. There are no error messages logged.

Error message

2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
46-282436P1 CARRIAGE TRAVEL SENSOR FLAG If the screws are broke or missing you will need two of them to hold the flag onto block. 46–265486P8 SCREW If the screws are broke or missing you will need two of them to hold the flag onto block. 46–265486P8 SCREW.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.