GE Healthcare · MRExciteDiscoveryOptima22493641081934956105992247156114092249317
CWA Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2249364 MGD Posix Sync Init could not find the shared data structure in the name database. 1081934956 3 1 Wed Apr 14 09:29:16 2004 2249364 4 CARD diagCoord mgdPosixSync.c 322 Task ID: 0x3f1fdf8 Time: 10599 Object: APS MGD Posix Sync Init could not find the shared data structure in the name database. vxWorks errno = 0x590001. 2247156 The IRF2 board was not detected in the MGD chassis Error: 2247156 Host: BOWIE-MRI Process: diagCoord File: Irf.cpp Line: 74 Task ID: 0x3f1fdf8 Time: 11409 Object: APS The IRF2 board was not detected in the MGD chassis. The system will not function without this board.-> 2249317 Communication problem between the APS, AGP, and SCP boards. MGD Resource Probe could not find the shared data structure in the name database. 1081934956 2 1 Wed Apr 14 09:29:16 2004 2249317 2 CARD diagCoord systemResourceApi.c 357 Task ID: 0x3f1fdf8 Time: 8585 Object: APS Communication problem between the APS, AGP, and SCP boards. MGD Resource Probe could not find the shared data structure in the name database. vxWorks errno = 0x590001.->
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.