GE Healthcare · MRExcite224858412078120702248592287272248579
SCIM communications errors due to defective WIM.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2248584 SCIM Communication Error. Thu Apr 10 07:21:09 2008 2248584 2 NSP SCP:KeyboardLinkHWO 1920 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications SCIM Communication Error. The SCIM failed to respond to a watchdog packet from the SCP. Watchdog packets waiting for a response: 0x0000 Check cable connection to Workspace Interface Module and to System Cabinet.-> EN 454 SR 436 1207812070 2248592 SCIM Configuration Failed SCIM link errors caused the configuration to fail 2248592 Host: PMR5_OC0 Process: NSP File: SCP:OpWorkspaceAutoConfig Line: 364 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Configuration Failed. SCIM link errors caused the configuration to fail. Please fix and perform a TPS Reset 2248579 A timeout occurred when trying to receive a SCIM packet 2248579 Host: PMR5_OC0 Process: NSP File: SCP:KeyboardLinkResources Line: 471 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Communication Error. A timeout occurred when trying to receive a SCIM packet. Opcode = 0x60 Check cable conn
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2141978-6 LIGHTNING WORKSPACE INTERFACE MODULE See long description 2141978-6 LIGHTNING WORKSPACE INTERFACE MODULE. If WIM is not available, order part 2141978-5 and replace cable between motherboard J1 and CIF J3. ** If replacing WIM with GOC AA, order cable 2395735.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.