Pular para o conteúdo principal
GE Healthcare · MROptimaBrivoExciteSigna

CWA System Applications Disk Is Full Due to usr Partition at 92 percent or more.

Symptom

Disk Space Problems See long description. LX systems will sometimes generate error messages indicating disk space problems. The message will recommend a system shutdown and reboot, which will run storelog to clear the problem. On occasion this will not work, and disk space limitations will still exist. There is a helpful file generated in these situations which can lead to the correct fix action. Check the /usr/g/service/log/diskfull.log file for LX disk space problems. LX systems will sometimes generate error messages indicating disk space problems. The message will recommend a system shutdown and reboot, which will run storelog to clear the problem. On occasion this will not work, and disk space limitations will still exist. There is a helpful file generated in these situations which can lead to the correct fix action. Check the /usr/g/service/log/diskfull.log file for LX disk space problems. {sdc@b3a}[1] cd /usr/g/service/log <enter> {sdc@b3a}[2] more diskfull.log <enter>

Error message

Application Disk Space Low, Removing Images Will Not Help Pop up Message. Application Disk Space Low, Removing Images Will Not Help Application Disk Space Low, Removing Images Will Not Help.

Applies to

Models
Optima, Brivo, Excite, Signa
Product
Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
Software
SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.