System will not reboot during software load.

Symptom

Software Fails to Boot and Crashes to the "OK" Prompt.

Error message

ID Prom invalid or The IDPROM contents are invalid Both messages appear in the boot up sequence, not in an error log. Won't see both at the same time. Message that is displayed depends on type of CPU. See Long Description. Partial example of the boot-up sequence for the AXI CPU; SPARCengine(tm)Ultra(tm) AXi (UltraSPARC-IIi 440MHz), No Keyboard OpenBoot 3.10.12 SME, 256 MB memory installed, Serial #303707648. Ethernet address 40:48:30:64:83:8b, Host ID: 921a3600. The IDPROM contents are invalid Boot device: disk3 File and args

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2210214 SUN ULTRA2 TIMEKEEPER PROM_SPARE Refer to Direction 2220946 Rev.13, Replacement/PM, OC, Host CPU, which shows the location of the NVRAM chip. Replace the chip as directed. Be sure to use appropriate static precautions. It is not necessary to remove the Video Board, just the Bit 3 (center) board to gain access to the chip.
Area
MR

Full cause & step-by-step resolution