Pular para o conteúdo principal
GE Healthcare · MRSignaExcite22581062258119

unable to auto configure a new Coil Due to Coil Can Not Be Auto Configured Because it is Not Compatible With Upgrade HDx…

unable to auto configure a new Coil Due to Coil Can Not Be Auto Configured Because it is Not Compatible With Upgrade HDxT, Only Forward Production HDxT.

Symptom

unable to auto configure a new Coil See long description Site is not able to auto configure a new coil when plugged into the coil port. System responds back with Do you want to configure the coil information? Site selects Yes, then the system states to install cd rom or cancel.

Error message

2258106 Coil Check Host coil validation error: Coil ID not found in CoilConfig.cfg file. Connector: 2258106 Host: MAINMR Process: coilCK File: coilCK_scp.c Line: 989 Exception Class: Unknown Function: Host: Coil Check Host coil validation error: Coil ID not found in CoilConfig.cfg file. Connector: Multicoil A Port Coil ID: SENTO1LA3LnrAinqRNcBLCl7dG5x0j4z.

2258119 A new coil is connected. This coil is not currently configured on your system. Do you want to configure the coil information? Mon Jan 19 11:14:05 2009 Error: 2258119 Host: HDe Process: coilCK File: coilCK_scp.c Line: 1565 Exception Class: Unknown Function: Host: Coil CheckA new coil is connected. This coil is not currently configured on your system. Do you want to configure the coil information?

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.