Table longitudinal drive does not drive in and out smoothly.
Symptom
Table longitudinal drive does not drive in and out smoothly. It will drive in or out a short distance, then stops See Long Description Table longitudinal drive does not drive in and out smoothly. It will drive in or out a short distance, then stop, and tech must release and hit the button again to get it moving. This happens with or without coils connected. But if the tech hits the Home button there is no delay at all, it goes right Home without stopping. There are no errors logged. Note that this is not related at all to the table drive workflow changes coming with HF03 M4+SP2 software as that FMI (FMI 67706) hasn't been installed on this system yet (and that FMI will not fix this problem). This is a system problem.
Error message
The Cradle is Obstructed This is a popup error message that the operator sees when the cradle is commanded to move and the system does not sense the movement. (The system counts encoder pulses per unit of time and will fail if none or too few pulses are seen). The error message is not logged in the message log.
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Component
- 2366642-4 CONNECTOR 37 CONTACTS 200 VOLTS 5 AMP See Long Description 2366642-4 CONNECTOR 37 CONTACTS 200 VOLTS 5 AMP Note: The part number for this connector is 2366642-4. There is an error in the part description for this part in FEMC where it is shown to be "CONNECTOR 9 CONTACTS 200 VOLTS 5 AMP", however it is actually a 37-pin connector. This was confirmed by the FE and by the engineering drawing in myWorkshop. The MR Program Integrator at headquarters was notified of this error on 09 Sep 2010 and is taking steps to fix this inaccurate part description in FEMC.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.