Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova105074311507431050734

Frontal & Lateral gantries are locked after system boot.

Symptom

INX1X1: Frontal & Lateral gantries are locked after system boot. Using the TSUI's, motion is not allowed when trying to move either frontal gantry or lateral gantry, whatever axis selected for motion. INX1X1: Frontal & Lateral gantries are locked after system boot. Using the TSUI's, motion is not allowed when trying to move either frontal gantry or lateral gantry, whatever axis selected for motion.

Error message

1050743 or 1150743 : Mismatch between MCB firmware and plane. See long description for further details. •Error code 1050743 & 1150743 - EM_MCB_FIRMWARE_INSTALL •Error text: Mismatch between MCB FirmWare and plane. Last download or card swap is incorrect •Class WARNING. •Display message: MCB FirmWare mismatch on Frontal or Lateral plane. Re-install. –Will remain displayed until cleared. –Display priority is the highest. •Reference to defective axis is also available through EM_FRT_XXXX_MCB_DOWNLOAD_FAILURE –1 error per axis, there is no ambiguity •Error is not triggered by empty/fresh MCB (i.e when MCB has not been loaded with firmware) This error was introduced with SPAD TR2 release which has 2 different MCB firmwares. The error informes at next boot if wrong firmware is used. Error 1050743 or 1150743 can be preceded by error 1050734 (Abort class) "POS : Incompatible "axis" MCB SW Version, that indicates in debug text the axis causing error

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Software
SPAD M4 TR2 MCB fix BIP_6.4.0, DL_SPAD_4.25.1 SPAD (Biplane) M4 TR2 MCB fix BIP_6.4.0, DL_SPAD_4.25.1 - MCB fix is related to frontal LC vibrations fix. SPAD (biplane) M4 TR2 MCB fix BIP_6.4.0, DL_SPAD_4.25.1. MCB fix is related to frontal LC vibrations fix.
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues