GE Healthcare · MRExciteSigna22246662244645
Gradient Processor Can Not Communicate With the Gradient Switch Box.
Symptom
Will Not Scan with Gradient Errors.
Error message
2224666 Auto Prescan failed, Master Gradient is not ready 2244645 - The GSC Stopped Responding to the Gradient Processor (GP). Thu Dec 11 09:31:21 2003 2244645 2 MR-OC0 NSP SCP:GP 471 The GSC stopped responding to the Gradient Processor (GP). Verify that the GSC is powered up and the serial communication cable is attached. If this does not work, try cycling GSC power.
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.