GE Healthcare · MRExcite22243471110252196
System intermittently locking up while scanning with Error 2224347 pixel header length field is inconsistent with recon…
System intermittently locking up while scanning with Error 2224347 pixel header length field is inconsistent with recon supplied data due to defective IT transfer card ( PCI ADAPTER CARD IT HOST )
Symptom
System intermittently locking up while scanning with Error 2224347 pixel header length field is inconsistent with recon supplied data .
Error message
2224347 pixel header length field (XXXX) is inconsistent with recon supplied data (xxxx) Error: 2224347 Host: lx-mr Process: TIR File: store_pixel_header.c Line: 340 pixel header length field (5204) is inconsistent with recon supplied data (1110252196)
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2235744-5 PCI ADAPTER CARD EXCITE II(IT HOST) also known as IT card Image Transfer
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.