Pular para o conteúdo principal
GE Healthcare · MRExciteSigna22522462254782

Prescan failures due to defective Receiver board in the DCERD2.

Symptom

Auto and Manual Prescan Fails

Error message

2252246 Quick reset detected on the CAN network. The node reset and recovered quickly. Recovered Node is Master DCERD Digital. 2252246 lx-mr NSP SCP:CanLinkHWO 776 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 Quick reset detected on the CAN network. The node reset and recovered quickly. Recovered Node is Master DCERD Digital.->

2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 NSP SCP:UPM 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. Peak value detected = 429 W. Max Peak Limit = 426 W- Possible cause(s): ASC RF Detector, ASC RF Processor, RF Amp, Software .

Applies to

Models
Excite, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5151531 DCERD2 4CH RCVR w_firmware
Area
MR

Full cause & step-by-step resolution