Vertical band artifacts on LAVA series using the 8ch body array coil due to bent pins on connector 2417162.

Symptom

MRIQ: Dark Vertical Bands or Venetian Blind Artifact in Images See attached file. The system is a 1.5T HiSpeed, with SGD Gradients, running 9.1 M4 software. These artifacts were pretty repeatable. The problem appeared to only occur on the Knee scans and some scans done with the TORSO coil. The Customer would see an artifact just about every other patient scan. There were intermittent errors in the log stating X axis overload and PPBM errors. See attached file. The system is a 1.5T HiSpeed, with SGD Gradients, running 9.1 M4 software. These artifacts were pretty repeatable. The problem appeared to only occur on the Knee scans and some scans done with the TORSO coil. The Customer would see an artifact just about every other patient scan. There were intermittent errors in the log stating X axis overload and PPBM errors.

Applies to

Models
Excite, HiSpeed, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
2417162 ASSY,FRU,CABLE,VALUE,BODY ARRAY (cable for coil M3335MC Signa 1.5T HD 8-Channel Body Array). See Long Description Note Even though the cable has "VALUE" in the part name, this is misleading because it isn't just used on HDe, it's also used on 1.5T HDxT, MR450, and MR450W as can be seen in the Coil Vendor Manuals sections of the respective Service Methods documentation as well as in the Coils Compatibility Matrix (Cable is a FRU for coil M3335MC Signa 1.5T HD 8-Channel Body Array). To view the most recent appropriate and detailed documentation, click to go to the Engineering Documents .
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues