Instrumentation Lead LEMO Connector is Loose Inside the Magnet.LCCI2 ALARM I2 HE LEVEL CBL OPEN.De-ice and Reseat Instru…

Instrumentation Lead LEMO Connector is Loose Inside the Magnet.LCCI2 ALARM I2 HE LEVEL CBL OPEN.De-ice and Reseat Instrumentation Lead LEMO Connector-wqr.Magnet LCC.LCCI2 ALARM I2 HE LEVEL CBL OPEN.

Symptom

LCCI2 ALARM I2 HE LEVEL CBL OPEN. LCCI2 ALARM I2 HE LEVEL CBL OPEN, the helium level cable has been disconnected or is damaged in a way that the Magnet Monitor 3 is detecting an open circuit.

Error message

LCCI2 ALARM I2 HE LEVEL CBL OPEN. LCCI2 ALARM I2 HE LEVEL CBL OPEN, the helium level cable has been disconnected or is damaged in a way that the Magnet Monitor 3 is detecting an open circuit.

Applies to

Product
Magnet LCC. Software Version Concept should NOT be selected for Magnet Products Serial #: 1.5T (Rxxx) / 1.0T (Qxxx) He Vessel Pressure:>3 psig but <5 psig Shield Temp:< 55 K Coldhead Temp:> 3.8K but < 4.5K Recondenser Temp:> 4.2K but < 4.55K Delta Temp (Coldhead - Recondensor):> 0 but < 0.3K Min. LHe Fill Level: 50% Min. LHe Ramp Level: 90% Min. LHe Shim Level: 75% Helium Volume:1940 L (1.5T) Helium Volume:1970 L (1.0T) Burst Disk <":2319285 Burst Disk >":2350415 Burst Disk O-ring:46-281101P1 Baffle Assy 2180818 Low Cryogen Consumption
Software
Magnet Monitor 3 Revision 1.56. Magnet Monitor 3 revision 1.56, distributed to sites with Connectivity automatically post FMI 60755 release on 6/23/2008. See Long Description for Fixes and Notes. Magnet Monitor 3 Revision 1.56 Software (formerly known as 1.55). Fixes New magnet type selection page with menus for both compressors. New remote commands to change the compressor selections. New uploaded properties to show the current compressor selections. A fix to the satellite interface so MM3 doesn't lose the sat link if it gets a malformed entitlement message. Changed configuration files so ABM and BOM magnets with DT500 sensors use the correct lookup tables. Disabled alarms actually won't generate alarms. The net mask no longer has to be entered with extra zeros (i.e. - 255.0.0.0 is now good enough instead of 255.000.000.000). The "Show Digital IO" web page shows all compressor types correctly - no more messages saying that Leybolds are inverted. New alarm limits for ABM and BOM for 2nd_stage_shield (is now 10K to 25K, was previously 3.5K to 10K). Changed Master Interval to 720 minutes for LCC and Non-Zero boil off magnets and 360 minutes for HFOs to reduce dispatch volume. NOTES •Software will be downloaded to all communicating sites except satellite modems. •Currently there is no CD that is orderable as a FRU for sites that do not have communication or have a satellite modem so that the software can be manually updated.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.