GE Healthcare · MRExcite2224842
Unexpected cradle motion Errors Due to BASIC SWITCH W/ OT PLUNGER in Dock.
Symptom
Table Movement is Rough, Jerks While Moving
Error message
2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 46-136334P56 BASIC SWITCH W/ OT PLUNGER Table Up Limit 46-136334P56 2 SWITCH 1 SPDT, BASIC SWITCH W/ OT PLUNGER.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.