GE Healthcare · X-ray VascularInnova36020863702086
169496 Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Symptom
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Error message
INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH. See long description for further details on Jedi err code description. INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH. Standard format message is: 3602086 (3702086) DO XRTechMgr Front (Lat): Jedi: LVPS problem Debug: jedi-error-code=0xxxh JEDI Error Code: 0xxx H " Error Text " Data Related to Error: 153 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: STAND_BY Error code and error text associated: 0531H - LVPS 400: heater 0 +160V too high 0532H - LVPS 400: heater 0 +160V too low 0533H - LVPS 400: heater 1 +160V too high 0534H - LVPS 400: heater 1 +160V too low 0535H - LVPS 400: main inverter +17V too high 0536H - LVPS 400: main inverter +17V too low 0537H - LVPS 400: main inverter +26V too high 0538H - LVPS 400: main inverter +26V too low 0539H - LVPS 400: main inverter +160V too high 0540H - LVPS 400: main inverter +160V too low
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- $ Unity M4 SP4, DL_Unity_4.19.14, FMI 12118 $ Unity M4 SP4, DL_Unity_4.19.14, FMI 12118 Date 01-01-2010 $ Unity M4 SP4, DL_Unity_4.19.14, FMI 12118 SP4 SYSTEM RELEASE UIS 4.52.1 DL GHOST 22.2.3 DL DL_Unity_4.19.14 GITANE V7.47 ITU OS: 2.2.4.6 RAFI: 1.11 GE: 1.3.3 Calypso CLY.PLATFORM 4.52.XX IDC 5.39.XX GITANE V9 FW1.5 POS CPU RR POS LBY 1.4.4U_23 PST 1.9.D PIB SW V3.6 FPGA V4.3 (with every type of interco lift) MCB DSP 34F / FPGA317 STEPPER BOARD IS6.31L JEDI P6.36BF1 AW AW4.4_04 or AW4.3_05.04 or AW4.2_07 3DXR 1.0.5.XX VOLUME VIEWER 6.5.0 or 6.7.0 or 6.8.3 or 7.4.71
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.