GE Healthcare · MRSignaExcite2252362
2252362 DD Driver Fault Detected Low Open Circuit Value due to Driver Module
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252362 DD Driver Fault Detected Low Open Circuit Value. Error: 2252362Host: SHMCMR01 Process: NSPFile: SCP:DCB Line: 1686DD Driver Fault Detected - Low Open Circuit Value.Possible loss of -15V supply at the DD output.Source of Fault is BD4 (Dynamic Disable Boards at Back of Body Coil).Value = 850 mV Limit: 1002 mV-
Applies to
- Models
- Signa, Excite
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.