GE Healthcare · MRExcite2244897
Equipment Room temperature exceeds GE specification window of 59-89.6°F (15-32 °C).
Symptom
Will Not Scan with Gradient Errors.
Error message
2244897 The GP is reporting a % Axis OverTemperature Fault See Long Description System will wait 5 minutes before attempting recovery. Extended Error: Check that the cabinet fan is functioning properly, rear door is closed and sealed, and that no filler panels have been removed from the front of the cabinet. Proper cooling requires a vaccum to pull air through the Gradient Amplifiers, Power Supply, and GP. If fault persists on specific PSDs write a PQR. If fault persists on many or all PSDs, replace Gradient Amplifier.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution