GE Healthcare · CTBrightSpeedVCT26011824326011801150075
BrightSpeed Select, Delight Select or MERI Select
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118243 (80-0280) Heater 0 board communication problem See LD for error explanation Error: 260118243 (80-0280) Heater 0 board communication problem Code 80 Hardware Errors Error code: 80-0280H Message/explanation: Heater 0 board communication problem PPC kV control board communication Watch Dog with Heater board popped up because it did not get reply from Heater board. Potential cause: Heater 0 board Control bus cable PPC kV control board Recommended action/ Troubleshooting guide: Verify CAN cable between LVPS board and Heater board is correctly connected Check that heater firmware is running (DS1-DS2 on heater board blinking). If firmware OK, replace PPC kV control board. Else verify that RESET Led is not lit. If it is lit, disconnect successively the control bus cable from Rotation and PPC kV control boards to find the board which is holding the reset line and replace it. If after disconnecting all the boards, the Led remains lit, replace Heater board Error: 260118011 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message See LD for complete error Fri Feb 3 07:59:35 2006 Error: 260118011 Host: orp Proces
Applies to
- Models
- BrightSpeed, VCT
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required