GE Healthcare · MRExcite22185372224842
Table moving off scan plane during scanning due to Table Position Encoder.
Symptom
Scan Pause occurs occasionally, must press Move to Scan button on the Keyboard to continue May occur when the cradle has to reposition itself axially during the scan per prescription.
Error message
2218537 Scan stopped, Scan Stopped due to error The table is not at scan plane See Long Description 2218537 MsaScanMgr.m 2852 Scan stopped, Scan Stopped due to error : The table is not at scan plane.. 2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec-
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5224109 Rotary Table Position Encoder 5224109 Rotary Table Position Encoder
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.