TPS failed to reset with Errors 2248102 2248128 and 2248011 due to APS board.

Symptom

System fails to boot Applications software and TPS reset fails. See long description. The system is a 1.5T LX, running 9.1 M4 software. The system will not boot applications. Looks like the Bit3 is down. Errors in the log look like TPS reset failure. Ping to tps fails as seen below: lx-mr 6# ping lx-mr-tps PING lx-mr-tps (216.33.27.2): 56 data bytes ping: sendto: Connection refused

Error message

2248102 software communications error. APS dacqControl received invalid cm packet %d, scan state = % Please perform a TPS Reset. 2248102 software communications error. APS dacqControl received invalid cm packet %d, scan state = % Please perform a TPS Reset.

2248011 The A/D Offsets have already been pushed. The values must be popped before another push is attempted Programmer or internal software error, please perform a TPS Reset 2248011 The A/D Offsets have already been pushed. The values must be popped before another push is attempted Programmer or internal software error, please perform a TPS Reset.

2248128 Software communications error APS DC cm_rev error 0 Please perform a TPS Reset

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2294300-4 APS BOARD WITH IT DAUGHTER BD
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues