GE Healthcare · MRSignaExcite2218537225235722522932252312

Low BD4 HV Bias in Body and Surface Modes due to Driver Module

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2218537 Scan stopped, Scan Stopped due to error. The error message key is - 2252357 Error: 2218537 Host: GVBC Process: scn File: MsaScanMgr.m Line: 2847 Scan stopped, Scan Stopped due to error. The error message key is - 2252357

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252312 DD Driver Fault Detected High Voltage Fault. Source of Fault is BD4 (Dynamic Disable Boards at Back of Body Coil). Error: 2252312 Host: MR3T Process: NSP File: SCP:DCB Line: 2024 DD Driver Fault Detected - High Voltage Fault. Source of Fault is BD4 (Dynamic Disable Boards at Back of Body Coil). Value = 392 V Limit: 425 V->

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.