GE Healthcare · MRExcite22494126087922521041273619358
After Replacing Damaged MGD Boards and Backplane MGD Diags Fail Due to Term Server Connection
Symptom
After Replacing Damaged MGD Boards and Backplane MGD Data Path Diags Fail After water damage FE replaced the MGD chassis, IRF I/O, STIF, IRF, DRF, and SCP, MGD data path diagnostics failed After water damage FE replaced the MGD chassis, IRF I/O, STIF, IRF, DRF, and SCP, MGD data path diagnostics failed.
Error message
2249412 A PCI error occured. The CPV3060 signalled a Master Abort Tue May 11 19:09:02 2010 Error: 2249412 Host: suf-mr1 Process: diagCoord File: excLog.c Line: 1357 Task ID: 0x3dcdef0 Time: 60879 Object: SCP A PCI error occured. The CPV3060 signalled a Master Abort. Qspan PCI Status Reg: 0x2290; Qspan Interrupt Status Reg: 0x10008C00 Qspan Interrupt Control Reg: 0x30600000. Additional information in /usr/g/service/log/vxPciLg-ecae 2252104 Executive Diagnostic software error. board SCP failed to read PPCbug booting pattern within 20 seconds. This error could be caused by board not responding to a reset command Tue May 11 19:09:18 2010 Error: 2252104 Host: suf-mr1 Process: diagCoord File: diags_utils.c Line: 336 Diagnostic: MOT Test Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1273619358 Object: Host Diag Executive Diagnostic software error. board SCP failed to read PPCbug booting pattern within 20 seconds. This error could be caused by board not responding to a reset command
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.