Some sites have reported corrupt SaveINFO data, especailly sites equipped with the Brainwave option.

Symptom

Corrupt SaveINFO data, especially sites equipped with the Brainwave option. See Long Description SMILe 3/10/2005 Subject : Prevent corruption of SaveINFO information saved to MOD. Effectivity: All MR Signa systems that perform a SaveINFO to backup configuration information. Problem: Some sites have reported corrupt SaveINFO data, especailly sites equipped with the Brainwave option. There is a known issue where saved protocols or paradigms (for Brainwave option) cannot have spaces in the name. Also illegal Unix characters (!@#$%^&*(){}[] <>?/ cannot be included in saved protocols or paradigms. Example for corrupted protocol: /usr/g/protocol/site/head/noggin scan This must be changed to /usr/g/protocol/site/head/nogginscan Example of corrupted Brainwave paradigm: /export/home/signa/brainwave/config/pdgm_instructions/right hand test This must be changed to /export/home/signa/brainwave/config/pdgm_instructions/righthandtest

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR