Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnovaAdvantx600104100001450127

Xray Abort Reposition Collimator Blades is dispayed on the Advantx console and Patient Dose Mismatch.

Symptom

IN2K: Xray Abort Reposition Collimator Blades is dispayed on the Advantx console and Patient Dose Mismatch. The operator indicates Xray Aborted Reposition Collimator Blades is dispayed on the Advantx console and Patient Dose Mismatch on the DL review monitor.

Error message

Innova: 600104 4F76 Collimator Can Heartbeat Error Innova Console: 4F76 Xray Abort Reposition coll blades. Patient Dose Mismatch.. on the SUIF Error Log: 600104 Collimator Control error Acquisition aborted; re-start. AR. Collimator CAN communication failure. 600104 AR XRBeamForm Collimator Control error Acquisition aborted; re-start Debug: XRBeamForm CAN Heart Beat Failure See LD for complete error. Innova Console: 4F76 Xray Abort Reposition coll blades. Patient Dose Mismatch.. on the SUIF Error Log: 600104 Collimator Control error Acquisition aborted; re-start. AR. Collimator CAN communication failure. 600104 AR XRBeamForm Collimator Control error Acquisition aborted; re-start Debug: XRBeamForm CAN Heart Beat Failure Check the power, Have found the 24v going to the footswitch to cause this issue by loading down the power supply..

error Code = 1000014, error Text = SID_SSD out of range, debug Text = SID out of range, recovery Class = IH, major Function /= XRImChainPos System impact: X-ray unavailable. Displayed message : "SID pb, use lift orange button to move detector to highest position". Note: This error is also logged when the positioner is resetted.

50127 POS : Call Service;

Applies to

Models
Innova, Advantx
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues