Pular para o conteúdo principal
GE Healthcare · MRExcite224459022445892244578

SGA Power Supply Internal Wiring Fault or Control Undervoltage Due to SGA Power Supply

Symptom

Will Not Scan with Gradient Errors.

Error message

2244590 The GP is reporting a SGA Power Supply OverTemperature Fault Error: 2244590 Host: MR04 Process: NSP File: SCP:Grad Line: 2975 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: IR Coil: HEAD Scan: 24/2 The GP is reporting a SGA Power Supply OverTemperature Fault. System will wait five minutes before attempting recovery.->

2244589 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage. 2244589 Host: tradmri01 Process: NSP File: SCP:Grad Line: 2903 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage.-

2244578 The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
SecondOpinion ACGD SGA Power Supply 2334862-2 This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution