Pular para o conteúdo principal
GE Healthcare · MRExcite22480082218537

2248008 RFM EOS semaphore timeout error Due to RRF backplane.

Symptom

Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14

Error message

2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.

2218537 Scan stopped, Scan Stopped due to error RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Wed Feb 10 15:57:13 2010 2218537 1 MR-HD scnMsaScanMgr.m 2852Scan stopped, Scan Stopped due to error : RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: %d FE ID: %d segCount: %d frameCount: %d.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2305884 Remote RF MOD ASM (Chassis w/midplane 2305884 Remote RF MOD ASM (Chassis w/midplane
Area
MR