Pular para o conteúdo principal
GE Healthcare · MRDiscoveryExciteOptima227570831260660224852422275562224666

CWA Unable to Download or Reset The TPS MGD or ISE.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2275708 The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. See Long Description Error: 2275708 Host: C-3TMRI2 Process: diagCoord File: RfHubCanHWO Line: 5792 Diagnostic: CFB Board Level Diagnostics Iteration: 1 of infinite Action: 1 Task ID: 0x2537bb0 Time: 31260660 Object: SCP Function: RF: RF Receive Communication Failure. The RF Hub failed to recover CAN communication on wakeup. The RRX is reporting that RF Hub power is OK and clock is on. Check RF Hub CAN cable connections.->

2248524 Message queue overflow. Perform a TPS Reset to recover. See Long Description 2248524JSMR1 NSPSCP:CanLinkHWO 1173Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1Message queue overflow. Perform a TPS Reset to recover.

2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. ***

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. 

Applies to

Models
Discovery, Excite, Optima
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.