Pular para o conteúdo principal
GE Healthcare · MRSignaExcite11093

Unable to scan with pop-up message "Prescription allocation request fail", due to disk partition "usr" at 100% from core…

Unable to scan with pop-up message "Prescription allocation request fail", due to disk partition "usr" at 100% from core.dcserver files created when invalid "Archive Device" was selected.

Symptom

Unable to scan with message Prescription allocation request fail System is able to prescan. But when the prescan ends and the scan is about to begin, it fails with a message "Prescription allocation request failed". This message doesn't appear in the log, but another one does (see below): Error: 2218079 Host: MR6 Process: msgView File: ImageStat.cc Line: 68 Image space query failed

Error message

Prescription Allocation Request Fail

11093 Failed to send images to be archived remotely. See Full Description for more details Jun 9 Wed 07:24:01 11093 dcserver dcscommon.c 244 Failed to send images to be archived remotely 1870/8 (CLRHMKIS02)

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues