Pular para o conteúdo principal
GE Healthcare · MRSignaExciteDiscoveryOptima22257022225693

CWA Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2225702 MDS link failure. The link appears to be broken between the IPG and the Gradient Subsystem. Fri Feb 18 05:47:41 2005 Error: 2225702 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_grad Line: 100 MDS link failure. The link appears to be broken between the IPG and the Gradient Subsystem. Check the connections and signal level of the MDS fiber optic link between the TYME II card and the Gradient Subsystem. Run MDS fiber optic link diagnostics.

2225693 An MDS device failed to respond to IPG. MDS opcode: 0x0A MDS link failure. An MDS device failed to respond to IPG. MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A Error: 2225693 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. Timer: 253 MDS header: 0x11 -(00/10 = Grad, 01/11 = RF Cabient) MDS opcode: 0x0A

Applies to

Models
Signa, Excite, Discovery, Optima
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required