GE Healthcare · MRExcite474733
Flip Angle on Spin Echo and Fast Spin Echo Series Changes to Zero After Transfer to PACS due to Non-GE PACS.
Symptom
Flip Angle on Spin Echo and Fast Spin Echo Series Changes to Zero After Transfer to PACS due to Non-GE PACS. See Long Description The "Flip Angle" shown on SE and FSE series changes from 90 to zero degrees after transfer to the customer's non-GE PACS. This is DICOM tag (0018,1314). There are no error messages. Reference PR # 474733.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution