GE Healthcare · MR
System Software Boot fails on SPARC20 Computer.
Symptom
System Software Boot fails on SPARC20 Computer Contour/ Profile with SPARC20 computer intermittently encounter boot failures from the OK prompt. The system then repeats the boot procedure without interaction from the operator.
Error message
Memory initialization failure This message appears during the boot process. It is not recorded in any error log.
Applies to
- Product
- Profile 1
- Software
- Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
- Component
- 2112412 64Mb Sun memory
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.