Pular para o conteúdo principal
GE Healthcare · MR

System Software Boot fails on SPARC20 Computer.

Symptom

System Software Boot fails on SPARC20 Computer Contour/ Profile with SPARC20 computer intermittently encounter boot failures from the OK prompt. The system then repeats the boot procedure without interaction from the operator.

Error message

Memory initialization failure This message appears during the boot process. It is not recorded in any error log.

Applies to

Product
Profile 1
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2112412 64Mb Sun memory
Area
MR