GE Healthcare · MR
What are the Safety Precautions for Ramped Magnets?
Symptom
What are the Safety Precautions for Ramped Magnets? Effectivity: All GE Superconducting Magnets. Introduction: A superconducting magnet is an energy storage device that can generate high internal voltages during a quench (rapid discharge of the field). The safety precautions identified below are required when servicing superconducting magnets. SMILe 2/3/2005
Applies to
- Product
- Magnet HFO. Software Version Concept should NOT be selected for Magnet Products Serial # (Txxx) Field strength: 0.7 T He Vessel Pressure >0.9 psig but <6.25 psig Shield Temp for Cryo-cooler coldhead < 40 K Temp for Shield-cooler Coldhead < 35 K Cryocooler 2nd stage Temp > 3.8K but < 4.5K Recondenser Temp > 4.2K but < 4.6K Delta Temp (Coldhead - Recondensor) < 0 but < 0.3K Min. LHe Fill Level 75% of total vessel Min. LHe Ramp Level 100% Helium Volume 340 Liters Burst Disk 2253802-2 Baffle Assy 2226764
- Area
- MR