Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.

Symptom

Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.

Error message

error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH, major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service" error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH , major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service"

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Detector conditioner 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172 5131740 SMC detector conditioner Detector conditioner Current rev is -15 but a -7 through -15 is acceptable. Detector Conditioner, connecter kit 2379172
Area
X-ray Vascular