GE Healthcare · MRExcite2254204
2254204 SRFD2 RF Amplifier Power Supply Fault due to SRFD2 RF amplifier.
Symptom
Stops Scanning with RF Amplifier Power Supply Faults Fri Jun 25 13:04:45 2004 Error: 2254204 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 1717 SRFD2 RF Amplifier Power Supply Fault. RF Amplifier indicates the high voltage power supply is out of specification.sible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated,supply FETS are shorted.-
Error message
2254204 SRFD2 RF Amplifier Power Supply Fault. RF Amplifier indicates the High Voltage Power Supply is out of Specification. Fri Jun 25 13:04:45 2004 Error: 2254204 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 1717 SRFD2 RF Amplifier Power Supply Fault. RF Amplifier indicates the high voltage power supply is out of specificati on.sible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated,supply FETS are shorted. Extended Error: Check fans, may be blocked or defective, Cabinet Filters need to be changed, or RF Amplifier may be operating out of specification.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.