Multi-Coil short error only when using multicoils due to internal short in SSM.
Symptom
Unable to Scan with Multi Coil TR Driver Faults
Error message
2225558 Multi-Coil T/R Driver Fault Multi-Coil Path Short Tue Jun 20 08:41:31 2006 Error: 2225558 Host: sommri Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6310 Multi-Coil T/R Driver Fault, Multi-Coil Path Short A short circuit was detected on port 4. Check cable to multi-coil board or check for a shorted pin diode.
Applies to
- Models
- Ovation, Excite, Signa
- Product
- Ovation with EXCITE
- Software
- OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
- Component
- 2160200-45 SSM assy for Ovation/OpenSpeed
- Area
- MR
Full cause & step-by-step resolution